Multizentrische Studie in der Zeitschrift Catheterization and Cardiovascular Interventions, 2024. Verfasst von Jorge Chavarria und 15 weiteren Autor:innen. Beteiligte Einrichtung: McMaster University and Hamilton Health Sciences, Hamilton, Canada. Hauptthemen laut MEDLINE: Registries, Transcatheter Aortic Valve Replacement, Aortic Valve, Prosthesis Design. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
The anatomic substrate of bicuspid valves may lead to suboptimal TAVR stent expansion and geometry.
Aim
We evaluated determinants of stent geometry in bicuspid valves treated with Sapien transcatheter aortic valve replacement (TAVR) valves.
Methods
A multicenter retrospective registry of patients (February 2019 to August 2022) who underwent post‐TAVR computed tomography to determine stent area (vs. nominal valve area) and stent ellipticity (maximum diameter/minimum diameter). Predictors of relative stent expansion (minimum area/average of inflow + outflow area) and stent ellipticity were evaluated in a multivariable regression model, including valve calcium volume (indexed by annular area), presence of raphe calcium, sinus diameters indexed by area‐derived annular diameter, and performance of pre‐dilation and post‐dilation.
Results
The registry enrolled 101 patients from four centers. The minimum stent area (vs. nominal area) was 88.1%, and the maximum ellipticity was 1.10, with both observed near the midframe of the valve in all cases. Relative stent expansion ≥90% was observed in 64/101 patients. The only significant predictor of relative stent expansion ≥90% was the performance of post‐dilation (OR: 4.79, p = 0.018). Relative stent expansion ≥90% was seen in 86% of patients with post‐dilation compared to 57% without ( p < 0.001). The stent ellipticity ≥1.1 was observed in 47/101 patients. The significant predictors of stent ellipticity ≥1.1 were the indexed maximum sinus diameter (OR: 0.582, p = 0.021) and indexed intercommisural diameter at 4 mm (OR: 2.42, p = 0.001). Stent expansion has a weak negative correlation with post‐TAVR mean gradient ( r = −0.324, p < 0.001).
Conclusion
Relative stent expansion ≥90% was associated with the performance of post‐dilation, and stent ellipticity ≥1.1 was associated with indexed intercommisural diameter and indexed maximum sinus diameter. Further studies to determine optimal deployment strategies in bicuspid valves are needed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- McMaster University and Hamilton Health Sciences, Hamilton, Canada.
- Ottawa Heart Institute, Ottawa, Canada.
- Centre for Cardiovascular Innovation, St. Paul's and Vancouver General Hospital, University of British Columbia, Vancouver, Canada.
- St. Michael's Hospital, University of Toronto, Toronto, Canada.
- Population Health Research Institute, Hamilton, Canada.
ORCID-Profile: Jorge Chavarria, Janar Sathananthan, Noman Ali
Zitieren
APA
Chavarria, J., Falcao, F., AlRaddadi, H., Aziz, A., Dick, A., Chung, K., Meier, D., Sathananthan, J., Ali, N., Um, K. J., Velianou, J., Natarajan, M., Jaffer, I., Wood, D., Fam, N., & Sheth, T. (2024). Bicuspid valve CT registry of balloon‐expandable TAVR: BETTER TAVR registry. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31091
Vancouver
Chavarria J, Falcao F, AlRaddadi H, Aziz A, Dick A, Chung K, et al. Bicuspid valve CT registry of balloon‐expandable TAVR: BETTER TAVR registry. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31091
BibTeX
@article{Chavarria2024,
author = {Chavarria, Jorge and Falcao, Felipe and AlRaddadi, Hatim and Aziz, Amir and Dick, Alexander and Chung, Kevin and Meier, David and Sathananthan, Janar and Ali, Noman and Um, Kevin John and Velianou, James and Natarajan, Madhu and Jaffer, Iqbal and Wood, David and Fam, Neil and Sheth, Tej},
title = {Bicuspid valve CT registry of balloon‐expandable TAVR: BETTER TAVR registry},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
doi = {10.1002/ccd.31091}
}
EUVIMED