EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Diffusion integrity testing of small‐area sterilizing‐grade membrane filters in single‐use assemblies

Anand Alembath iD, Joseph Hersey iD

Biotechnology Progress · 10.11.2026

DOI: 10.1002/btpr.88563 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biotechnology Progress, online erschienen am 10.11.2026. Verfasst von Anand Alembath und 1 weiteren Autor:innen. Beteiligte Einrichtung: MilliporeSigma, the U.S. and Canada Life Sciences Business of Merck KGaA, Darmstadt, Germany. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Sterile filtration represents the final critical step in biomanufacturing, where a sterilizing‐grade membrane filter removes microbial contamination from the drug product — making filter integrity verification essential. While bubble point and gas–liquid diffusion tests are established non‐destructive methods for hydrophilic membranes, diffusion testing offers high sensitivity to small defects and operates at lower test pressures. However, applying diffusion testing to small‐area filters integrated into closed single‐use assemblies remains challenging because pressure changes measured by automated integrity testers can arise not only from membrane diffusion, but also from assembly‐related effects such as gas‐permeable tubing, upstream volume and thermal equilibration. For low‐area devices, these background contributions can become comparable to or greater than the membrane signal, reducing test sensitivity and complicating interpretation. This work presents a practical method for implementing diffusion‐based integrity testing of small‐format sterilizing‐grade filters in final sealed single‐use assemblies. The approach quantifies assembly‐specific background contribution and incorporates it into an adjusted diffusion specification, allowing membrane‐specific diffusion to be resolved without altering the assembled system. By accounting for system‐derived measurement noise, the method enables reliable end‐user integrity assessment of closed assemblies that cannot be reconfigured to reduce upstream volume or replace high‐permeability tubing materials.

Schlagwörter

Stichwörter der Autor:innen

automated integrity testersbiomanufacturingclosed systemdiffusion testfilter integritysingle‐use assemblysterile filtration

Autor:innen und Institutionen

  • MilliporeSigma, the U.S. and Canada Life Sciences Business of Merck KGaA, Darmstadt, Germany.

ORCID-Profile: Anand Alembath, Joseph Hersey

Zitieren

APA

Alembath, A., & Hersey, J. (2026). Diffusion integrity testing of small‐area sterilizing‐grade membrane filters in single‐use assemblies. Biotechnology Progress. https://doi.org/10.1002/btpr.88563

Vancouver

Alembath A, Hersey J. Diffusion integrity testing of small‐area sterilizing‐grade membrane filters in single‐use assemblies. Biotechnology Progress. 2026. doi:10.1002/btpr.88563

BibTeX
@article{Alembath2026,
  author = {Alembath, Anand and Hersey, Joseph},
  title = {Diffusion integrity testing of small‐area sterilizing‐grade membrane filters in single‐use assemblies},
  journal = {Biotechnology Progress},
  year = {2026},
  doi = {10.1002/btpr.88563}
}
Kostenlos starten →