Commissioning and performance evaluations of the electron Monte Carlo algorithm in a new commercial treatment planning system

Feng Wang iD, Xuming Jiang, Zhi Shi, Yanfang Liu, Yibin Zhang iD, Jiayi Chen

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70698 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Feng Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Monte Carlo Method, Algorithms, Electrons. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Noncommunicable Chronic Diseases-National Science and Technology Major Project, Shanghai Innovative Medical Device Application Demonstration Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study provides a comprehensive evaluation of the newly implemented electron Monte Carlo dose calculation algorithm in the commercial treatment planning system, uTPS. It investigates the performance of the algorithm by examining its dosimetric accuracy and calculation speed, which have not been previously reported in the literature.

Methods

To evaluate the dose calculation accuracy of the electron Monte Carlo algorithm in uTPS, the beam data from the electron energies of 6 MeV, 9 MeV, and 12 MeV in the newly developed dual‐layer MLC CT‐linac were modeled. Following AAPM MPPG 5.b, the calculated point/plane doses under normal and extended SSD, oblique incidence and cutouts, and heterogeneous phantoms with different energies were compared with the corresponding measurement data. The calculation time was evaluated by the comparisons between two different calculation workstations.

Results

The beam data of 6 MeV, 9 MeV, and 12 MeV are well reproduced by the calculations. A dose grid resolution of 2 mm, 1% uncertainty, and dose to medium mode were used. Calculations successfully validated against the measured beam data, achieving mean dose differences of less than 2% in a water tank. The gamma passing rates were generally high and mostly exceeded 95% using 3 mm/3% criteria, although several larger‐field, extended‐SSD, and oblique/cutout cases showed lower passing rates, with minimum values remaining above 90%. The cutout tests, both special designed and patient specific, demonstrate that the algorithm can handle cases with cutouts effectively. In the heterogeneous phantom tests, the point dose differences for all SSDs are within 3.3%, indicating the accuracy of the uTPS Monte Carlo algorithm in handling heterogeneous media. The calculation times ranged from approximately 5 – 30 s for a statistical uncertainty of 1%, whereas achieving a tighter uncertainty of 0.5% increased the computation time to between 10 s and 2.5 min.

Conclusion

Verifications performed under these basic and complicated conditions indicate that the accuracy of electron Monte Carlo algorithm in uTPS meets the requirements of MPPG 5.b for electron beam dose verification. The electron Monte Carlo algorithm of uTPS demonstrates accuracy comparable to that of other commercial TPSs. The comprehensive evaluation confirms the algorithm's robust performance, as it achieves the necessary levels of dosimetric precision and calculation times to support diverse radiotherapy workflows.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedMonte Carlo MethodAlgorithmsElectronsPhantoms, ImagingParticle AcceleratorsNeoplasmsRadiotherapy DosageHumansRadiotherapy, Intensity-ModulatedRadiometry

Stichwörter der Autor:innen

and uTPSdose calculation accuracydose calculation timeelectron monte carlo algorithm

Autor:innen und Institutionen

  • Department of Radiation Oncology, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
  • Shanghai Key Laboratory of Proton Therapy, Shanghai, China.
  • Institute for Medical Imaging Technology, Shanghai, China.
  • Shanghai United Imaging Healthcare Co., Shanghai, China.

ORCID-Profile: Feng Wang, Yibin Zhang

Förderung

  • National Natural Science Foundation of China
  • Noncommunicable Chronic Diseases-National Science and Technology Major Project
  • Shanghai Innovative Medical Device Application Demonstration Project

Zitieren

APA

Wang, F., Jiang, X., Shi, Z., Liu, Y., Zhang, Y., & Chen, J. (2026). Commissioning and performance evaluations of the electron Monte Carlo algorithm in a new commercial treatment planning system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70698

Vancouver

Wang F, Jiang X, Shi Z, Liu Y, Zhang Y, Chen J. Commissioning and performance evaluations of the electron Monte Carlo algorithm in a new commercial treatment planning system. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70698

BibTeX
@article{Wang2026,
  author = {Wang, Feng and Jiang, Xuming and Shi, Zhi and Liu, Yanfang and Zhang, Yibin and Chen, Jiayi},
  title = {Commissioning and performance evaluations of the electron Monte Carlo algorithm in a new commercial treatment planning system},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70698}
}
Kostenlos starten →