EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Commissioning and clinical implementation of Monaco treatment planning system on a compact pencil beam scanning proton therapy system

Xiangkun Xu, Xiaoda Cong, Jian Liang, An Qin iD, Shupeng Chen, Xiaoqiang Li, Weili Zheng, Peilin Liu iD, Raymond Dalfsen, James Dolan, Martin Soukup, Craig Stevens und 2 weitere

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70587 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Xiangkun Xu und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Corewell Health William Beaumont University Hospital, Royal Oak, Michigan, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Proton Therapy, Monte Carlo Method, Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, Health Foundation Southeast Michigan, NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Treatment planning system (TPS) requires comprehensive commissioning and validation prior to clinical use to ensure accurate dose calculation and safe clinical implementation.

Purpose

To enable clinical use of the Monaco proton TPS, we commissioned it on a single‐room compact proton therapy system and implemented a clinically integrated workflow using daily synthetic CT (sCT) to track fractional dose delivery.

Methods

The Monte Carlo (MC) dose calculation algorithm in Monaco TPS (version 6.1) was commissioned on an IBA Proteus ONE system. Commissioning included data acquisition, beam modeling, and model validation. Data included in‐air spot profiles for 70–227 MeV at five planes (+20 cm and +10 cm upstream, isocenter, −10 cm and −20 cm downstream), integral depth‐dose curves in water, and absolute monitor unit calibration. Model validation comprised: percent depth‐dose in water for monoenergetic beams using a 10 × 10 cm 2 field across energies; output checks with standard cube plans mimicking diverse treatment sites and target volumes; and independent end‐to‐end testing with an IROC brain phantom. Clinically, a patient with basal cell carcinoma of the skin was treated with a two‐field IMPT plan delivering 60 Gy in 30 fractions. Probability‐based robust optimization required 100% target coverage under 3 mm setup and 3.5% range uncertainties. An in‐house sCT generated from daily CBCT images was used to evaluate dose delivery accuracy throughout treatment.

Results

Distal range (R90, R80, R20) differences between Monaco and measurement were within 1 mm for 70–227.7 MeV. Output differences for all standard cube plans were within 3%. For the IROC brain phantom, the TLD‐to‐Monaco MC dose ratio was 1.00, with gamma passing rates of 94% (coronal) and 98% (sagittal). sCT‐based evaluation showed 99% CTV receiving ≥98% of the prescribed dose across the course.

Conclusions

Monaco's proton MC beam model and dose calculation algorithm were commissioned and validated on IBA Proteus ONE for clinical use. The first patient (November, 2024) was successfully treated, with daily dose delivery accuracy monitored via an in‐house sCT platform.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedProton TherapyMonte Carlo MethodAlgorithmsPhantoms, ImagingSkin NeoplasmsHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedOrgans at Risk

Stichwörter der Autor:innen

Treatment planning systemcommissioningproton therapy

Autor:innen und Institutionen

  • Department of Radiation Oncology, Corewell Health William Beaumont University Hospital, Royal Oak, Michigan, USA.
  • Elekta AB, Stockholm, Sweden.

ORCID-Profile: An Qin, Peilin Liu

Förderung

  • NCI NIH HHS
  • Health Foundation Southeast Michigan
  • NIH HHS
  • Elekta and IBA
  • National Science Foundation

Zitieren

APA

Xu, X., Cong, X., Liang, J., Qin, A., Chen, S., Li, X., Zheng, W., Liu, P., Dalfsen, R., Dolan, J., Soukup, M., Stevens, C., Deraniyagala, R., & Ding, X. (2026). Commissioning and clinical implementation of Monaco treatment planning system on a compact pencil beam scanning proton therapy system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70587

Vancouver

Xu X, Cong X, Liang J, Qin A, Chen S, Li X, et al. Commissioning and clinical implementation of Monaco treatment planning system on a compact pencil beam scanning proton therapy system. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70587

BibTeX
@article{Xu2026,
  author = {Xu, Xiangkun and Cong, Xiaoda and Liang, Jian and Qin, An and Chen, Shupeng and Li, Xiaoqiang and Zheng, Weili and Liu, Peilin and Dalfsen, Raymond and Dolan, James and Soukup, Martin and Stevens, Craig and Deraniyagala, Rohan and Ding, Xuanfeng},
  title = {Commissioning and clinical implementation of Monaco treatment planning system on a compact pencil beam scanning proton therapy system},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70587}
}
Kostenlos starten →